Detail Multi-Echo Ultrasonic Thickness Gauge
Testing Instrument
Multi-Echo Ultrasonic Thickness Gauge
The LM series ultrasonic thickness gauge can measure the thickness of metal materials, high-temperature pipelines, pipes, plates (steel plates, aluminum plates), plastics, iron pipes, PVC pipes, glass and other materials. Can be applied to the detection fields of manufacturing industry, metal processing industry, chemical industry, commodity inspection industry and the like, can also monitor various pipelines and pressure vessels in production equipment and monitor the thinning degree of the pipelines and pressure vessels after being corroded in the use process, and is a corresponding instrument for material protection.
DESCRIPTION
Our company has introduced a hand-held precision ultrasonic thickness gauge LUM-3 using single crystal delay line probe, which uses single crystal delay probe and uses multiple echoes to improve accuracy, so that the resolution can reach 0.001mm and the lower limit of measurement can reach 0.30 mm. LM-3 has two working modes: normal and precise. The common mode adopts the interface wave-echo method, and the measuring range is 1.5 mm ~ 20 mm; The precision mode adopts echo-echo method, and the measurement range is 0.30mm ~ 10 mm.
Suitable for measuring flat sheet steel parts.
FEATURES
1. High precision, 0.001mm resolution
2. Suitable for measuring thin materials, can accurately and reliably measure 0.25mm flat thin steel plate material
3. Can penetrate through the coating, such as paint layer on the material, without polishing, can directly measure the thickness of the substrate
SPECIFICATIONS
Model
LM-3
Display
128× 64 Dot Matrix LCD Display (EL Backlight)
Working principle
Interface wave-echo mode (normal mode)
echo-echo mode (precision mode)
Measuring range
1.5~20mm (Normal mode) 0.3~10mm (Precision mode)
Unit and Display Resolution
Millimeter-0.001,0.01,0.1 inch-0.0001,0.001,0.01
Indication error
±0.01 mm (below 3mm) ;±0.05 mm (below 20 mm)
Calibration mode
One-point calibration, two-point calibration
The measurement update rate
For routine measurement: 4Hz;
scanning the minimum value:25 Hz
The sound speed range of the material
1000~9999m/s or 0.0394~0.3937in/us
The storage function
divided into 5 arrays, which can store 500 thickness values.
Sound speed storage
store the sound speed of 5 different materials.
Language
Chinese or English
Power supply
two 1.5V AA batteries. When the power is insufficient, there will be a low voltage prompt.
Operating time
alkaline battery can be used for 200 hours (no backlight)
Automatically shut down
after 5 minutes without operation
Working temperature
-10℃ ~+50℃, with special requirements up to -20℃
Size
149*73*32mm
Weight
200g (including battery)