Detail DW-F98 Fluorescence Spectrophotometer Brand Drawell

Spectrometer

DW-F98 Fluorescence Spectrophotometer
DW-F98 fluorescence spectrophotometer is a new generation of high-performance molecular luminescence analysis equipment. This product is designed for high performance, with high signal to noise ratio, ultra-high scanning speed, ultra-high resolution, ultra-high wavelength accuracy and a variety of accessories. This instrument easily meets the requirements in material research, drug analysis, biochemical and clinical testing, water quality control, food safety testing and other areas of qualitative and quantitative analysis and scientific research. Quantum yield accessory is available for F98 fluorescence spectrophotometer.
Main Features of DW-F98 Fluorescence Spectrophotometer
Horizontal slit optical design, with excellent luminous detection efficiency and high signal to noise ratio. Water Raman peak SNR better than 350: 1 (P-P) or 1500: 1 (RMS). The minimum sample volume is 0.5mL when using a standard 10mm square sample cell.
Wavelength scanning speed up to 60000 nm / min.
Bandwidth 1/2/5/10/20nm adjustable.
Fluorescence Quantum Yield accessory available.
The software offers fluorescence 3D scanning, equal-wavelength difference synchronous scanning, equal wave number difference (constant energy difference) synchronous scanning functions. (synchronous scanning function require pro version software)
Multi accessories available.
Using Hamamatsu’s high quality Xenon light source and photoelectric multiplier tube detectors and instruments to provide sufficient light intensity signal and the detection sensitivity.
Built-in frequency filter. Built-in optical gate, designed for unstable sample

Home / All products / DW-F98 Fluorescence Spectrophotometer













DW-F98 Fluorescence Spectrophotometer
DW-F98 fluorescence spectrophotometer is a new generation of high-performance molecular luminescence analysis equipment. This product is designed for high performance, with high signal to noise ratio, ultra-high scanning speed, ultra-high resolution, ultra-high wavelength accuracy and a variety of accessories. This instrument easily meets the requirements in material research, drug analysis, biochemical and clinical testing, water quality control, food safety testing and other areas of qualitative and quantitative analysis and scientific research. Quantum yield accessory is available for F98 fluorescence spectrophotometer.

DW-F98 Manual
Get Free Quote
Description
Main Features of DW-F98 Fluorescence Spectrophotometer
Horizontal slit optical design, with excellent luminous detection efficiency and high signal to noise ratio. Water Raman peak SNR better than 350: 1 (P-P) or 1500: 1 (RMS). The minimum sample volume is 0.5mL when using a standard 10mm square sample cell.
Wavelength scanning speed up to 60000 nm / min.
Bandwidth 1/2/5/10/20nm adjustable.
Fluorescence Quantum Yield accessory available.
The software offers fluorescence 3D scanning, equal-wavelength difference synchronous scanning, equal wave number difference (constant energy difference) synchronous scanning functions. (synchronous scanning function require pro version software)
Multi accessories available.
Using Hamamatsu’s high quality Xenon light source and photoelectric multiplier tube detectors and instruments to provide sufficient light intensity signal and the detection sensitivity.
Built-in frequency filter. Built-in optical gate, designed for unstable sample.


Specifications of DW-F98 Fluorescence Spectrophotometer
Product Name
F98 Series Fluorescence Spectrophotometer
Picture

Excitation Source
150W xenon lamp(Hamamatsu)
Excitation Wavelength
200nm~900nm
Emission Wavelength
200nm~900nm
Excitation Slit
1nm/2nm/5nm/10nm/20nm
Emission Slit
1nm/2nm/5nm/10nm/20nm
Wavelength Accuracy
±0.4nm
Wavelength Repeatability
≤0.2nm
Signal-to-Noise Ratio
S/N≥350(P-P) S/N≥1500(RMS) S/N≥15000(RMS BG)
Limit
≤5×10-11 g/ml(Quinine Sulfate Solution)
Linearity
γ≥0.995
Peak Repeatability
≤1.5%
Wavelength Scan Speed
Multi-speed Level, Maximum at 60000nm/min
Minimum sample size
0.5mL (Using a standard 10mm square sample cell)
Octave filter
Build in Octave filter
Photometric Quantity Range
-9999 ~ 9999
Response Time
0.02/0.1/0.5/1/2/4/8 s,Auto adjust
Data Transportation
USB2.0
Power
200W
Power Source
AC 220V/50Hz; 110V/60Hz
Instrument Dimension
610×460×365(mm)
Weight
Net Weight:21kg Gross Weight:26kg