Detail Bante904P Portable Conductivity/DO Meter
Alat Laboratorium Air
Bante904P Portable Conductivity/DO Meter
Multiparameter portable conductivity/DO meter contains 5 measurement modes, the meter is suitable for measuring the conductivity, TDS, salinity, resistivity and DO.
Conductivity/TDS/Salinity/Resistivity Modes:
• Multiparameter portable conductivity/DO meter is equipped with a backlit LCD display.
• 1 to 5 points calibration with automatic recognition for conductivity standards.
• Selectable cell constant (K=0.1/1/10), reference temperature (20/25°C), TDS factor, linear
and pure water compensations, seawater and practical salinity measurement modes.
Dissolved Oxygen Mode:
• 1 or 2 points calibration using the air-saturated water or zero oxygen solution.
• Salinity and barometric pressure compensations eliminate the measurement error.
Other Features:
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-Read function senses and locks the measurement endpoint.
• Calibration due alarm prompts user to calibrate the meter regularly.
• Auto-Power Off effectively conserves battery life.
• Setup menu allows setting the number of calibration points, stability criteria, temperature
unit, date and time, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• USB communication interface is easy to transfer data to PC.
• Multi-mode power scheme (batteries, power adapter and USB port) ensures that using the
meter smoothly.
SpecificationsOrdering InfoOptional AccessoriesPackaging
Model
Bante904P
Conductivity
Range
0.01 to 20.00, 200.0, 2000µS/cm, 20.00, 200.0mS/cm
Resolution
0.001, 0.01, 0.1, 1
Accuracy
±0.5% F.S.
Calibration Points
1 to 5 points
Calibration Solutions
10µS, 84µS, 1413µS, 12.88mS, 111.8mS
Temperature Coefficient
Linear (0.0 to 10.0%/°C), pure water
Reference Temperature
20/25°C
Cell Constant
K=0.1, 1, 10 or custom
TDS
Range
0.00 to 10.00, 100.0, 1000ppm, 10.00, 200.0ppt
Resolution
0.01, 0.1, 1
Accuracy
±1% F.S.
TDS Factor
0.1 to 1.0 (default 0.5)
Salinity
Range
0.00 to 80.00ppt, 0.00 to 42.00psu
Resolution
0.01
Accuracy
±1% F.S.
Resistivity
Range
0.00 to 20.00MΩ
Resolution
0.01
Accuracy
±1% F.S.
Dissolved Oxygen
DO Range
0.00 to 20.00mg/L
Resolution
0.01, 0.1mg/L, selectable
Accuracy
±0.2mg/L
% Saturation of Oxygen
0.0 to 200.0%
Resolution
0.1, 1%, selectable
Accuracy
±2.0%
Calibration Points
1 or 2 points
Pressure Correction
60.0 to 112.5kPa/450 to 850mmHg, manual
Salinity Correction
0.0 to 50.0g/L, manual
Temperature
Range
0 to 105°C/32 to 221°F
Resolution
0.1°C/0.1°F
Accuracy
±0.5°C/±0.9°F
Offset Calibration
1 point
Calibration Range
Reading ±10°C
General Specifications
Temperature Compensation
0 to 100°C/32 to 212°F, manual or automatic
Stability Criteria
Low or high
Calibration Due Alarm
1 to 31 days or off
Auto-Power Off
10, 20 or 30 minutes after last key pressed
Memory
Stores up to 500 data sets
Output
USB communication interface
Connector
6-pin nimi-DIN, 3.5mm jack socket
Display
Custom LCD (80×60mm)
Power
3×1.5V AA batteries or DC5V power adapter
Battery Life
Approximately 150 hours (turn off the backlight)
Dimensions
170(L)×85(W)×30(H)mm
Weight
300g