Professional laboratory pH/ion meter contains 5 ion concentration measurement methods and water hardness measurement mode, 1 to 5 points calibration, accuracy: 0.002pH.
pH Mode:
• Professional laboratory pH/ion meter is equipped with a 7 inch TFT display.
• 1 to 5 points calibration with automatic recognition for USA, NIST and DIN buffers.
• Solution Temperature Coefficient compensates for the pure water samples and references
the pH to 25°C.
ORP Mode:
• 1 point offset calibration allows adjusting the displayed value to a known standard.
• Relative and absolute millivolt modes ensure the reliable ORP measurements.
Ion Concentration Mode:
• 2 to 5 points calibration, including the 8 concentration points can be selected.
• Electrode management is capable of storing and recalling up to 3 electrode slopes.
• Selectable ion measurement methods (direct reading, known addition, known subtraction,
sample addition, sample subtraction) and concentration units (ppm, mg/L, mol/L, mmol/L).
Water Hardness Mode:
• 2 to 5 points calibration from low to high concentrations.
• Selectable measurement units (german degree, english degree, french degree, mmol/L
and mg/L) are used for professional water hardness measurements.
Other Features:
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-Read function senses and locks the measurement endpoint.
• Timed Interval Readings sends the measurement data to a PC or printer.
• Limit Alarm automatically alerts when reading exceeds the specified range.
• Calibration due alarm prompts user to calibrate the meter regularly.
• Calibration report shows the date, time, calibration points, slopes and offset.
• Password protection prevents the unauthorized calibration and settings.
• Expanded memory stores or recalls up to 1000 data sets.
• Reset function automatically resumes all settings back to the factory defaults.
Model
A130
pH
Range
-2.000 to 20.000pH
Resolution
0.001, 0.01pH, selectable
Accuracy
±0.002pH
Calibration Points
1 to 5 points
pH Buffer Options
USA, NIST, DIN, 5 custom buffers
ORP
Range
±2000.0mV
Resolution
0.1mV
Accuracy
±0.2mV
Calibration Points
1 point
Measurement Modes
Relative or absolute mV
Ion Concentration
Range
0.001 to 30000 (deponding on the range of ISE)
Resolution
0.001, 0.01, 0.1, 1
Accuracy
±0.5% F.S. (monovalent), ±1% F.S. (divalent)
Calibration Points
2 to 5 points
Calibration Solutions
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000
Measurement Units
ppm, mg/L, mol/L, mmol/L
Measurement Methods
Direct reading, known addition, known subtraction,
sample addition, sample subtraction
Water Hardness
Range (Concentration)
0.05 to 200mmol/L
Range (German Degree)
0 to 1120°dH
Range (English Degree)
0 to 1404°e
Range (French Degree)
0 to 2000°fH
Range (Ca2+)
0 to 8000mg/L
Range (CaCO3)
0 to 19999mg/L
Range (CaO)
Resolution
0.001, 0.01, 0.1, 1
Accuracy
±1% F.S.
Calibration Points
2 to 5 points
Calibration Solutions
0.01, 0.1, 1, 10, 100mmol/L
Temperature
Range
0 to 105°C/32 to 221°F
Resolution
0.1°C/0.1°F
Accuracy
±0.5°C/±0.9°F
Offset Calibration
1 point
Calibration Range
Reading ±10°C
General Specifications
Temperature Compensation
0 to 100°C/32 to 212°F, manual or automatic
Stability Criteria
Standard or high-accuracy
Calibration Due Alarm
1 to 31 days or off
Interval Readings
10, 30, 60 seconds, 10, 30 minutes or off
Password Protection
4 digits
Memory
Stores up to 1000 data sets
Output
USB communication interface
Connector
BNC, 3.5mm jack socket
Display
7” TFT LCD
Power
DC12V, using AC adapter, 220V/50Hz
Dimensions
240(L)×220(W)×80(H)mm
Weight
1.7kg
0 to 11220mg/L